Product Code: 2389
The KEP AMP-1 signal conditioner amplifies the low level signals from a magnetic pickup or flow transmitter by a factor of more than a hundred times
to drive any ratemeter, counter or controller. The unit uses only 30 mA and operates from 11 to 26 VDC. It has a 2.7K pullup resistor
attached to the open collector output and sinks a minimum of 100 mA to less than 1 V from a maximum of 26 VDC. It is mounted
in a rugged 2″ x 3″ metal housing with screw terminals for easy installation. Operating temperature is 32 to 140°F (0 to 60°C). Note: The low voltage line from the magnetic pickup to the AMP-1 should be less than 10 ft. in length, shielded and isolated from
relays, solenoids or other sources of electrical noise (let the output line make the long run). If the input is too sensitive, lower the 1.1K
input impedance by adding a 220 to 1K ohm resistor across TB1, pins 1 and 2, to increase noise immunity. A special version, the AMP1-10k, is available with 10 kΩ impedance for use with turbine and paddle wheel flowmeters.
Megger TTRU1 Handheld Transformer Turns Ratiometer
E+H Liquiline M CM42 Transmitter
Waltron 3151 Hardness Analyzer
Extech 461920 Mini Laser Photo Tachometer Counter
Fluke 1662 Multifunction Installation Tester
Rosemount 8800D Series Vortex Flow Meter
Fluke 568 Ex Intrinsically Safe Infrared Thermometer
E+E Omniport 30 Multi Function Monitor
Michell Instruments PCMini52 Humidity Mini Probe
Watlow EZ-ZONE RM Control Module (RMC) Multi-Function Controller
Kanomax 6006 Anemomaster LITE Anemometer
King Instrument 7100 Series Rotameter
Extech 475040 and 475044 Digital Force Gauges
Time Electronics 1010 DC Voltage Calibrator
Watlow MI Band Heater
Extech 461825 Photo Tachometer / Stroboscope